Rally #45-1-1185
Baxter/Hillrom/Welch Allyn #VS100S-B
Welch Allyn Spot Vision Screener (1/EA)
Welch Allyn VS100 Spot Vision Screener with Carrying Case and 8' (2.5m) US Power Cord (2.5A, 125V).
Features
- A handheld, portable ocular screening device designed to identify risk factors of amblyopia vision issues in patients as young as six months old.
- Evaluates light reflexes from the retina to estimate refractive error, pupil size, pupil distance, and eye gaze deviation.
- Automatically captures images from a non-invasive, three-foot distance using fixation target and sounds.
- Displays results quickly, indicating whether measurements are in range or a complete eye exam is recommended.
- EMR integration available for Epic and Cerner with the SpotConnect EMR application.
- Captures readings 97% of the time to help screen otherwise difficult subjects.
- Built-in driverless printing for efficient screening experience.
- Summary reports automatically uploaded with SpotConnect, exported wirelessly or via USB.
Important Notes:
- Rx Only
- Product is non-returnable; please make sure it is the right product before ordering.
- Warning: Product is considered HAZMAT and cannot ship via Air or to AK, GU, HI, PR, VI.
Warning: This product can expose you to chemicals known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.
Rally # | 45-1-1185 |
Manufacturer # | VS100S-B |
GTIN | 00732094212228 |
Brand | Welch Allyn® Spot® |
Manufacturer | Baxter / Hillrom / Welch Allyn |
Country of Origin | Unknown |
Application | Eye Exam Instrument |
Instrument | Type |
Dimensions | 8 1/2 in x 6 3/4 in x 4 3/4 in (21.6 cm x 17.1 cm x 12.1 cm) |
Weight | ~2.55 lbs (1157 g) |
Wireless Network | 802.11 a/b/g/n |
Expected Battery Life | 2.5 years (under normal usage) |
UNSPSC Code | 42183034 |
Warranty | One Year Warranty |
Item | Specification |
MFG | FRIWO |
Model | FW8030M/DT/15 |
Input Voltage | 90 – 264VAC (100 – 240VAC Nominal) +/- 10% |
Input Frequency | 47 – 63 Hz |
Input Current | 0.6-0.3Arms @ Max Load |
Avg. Efficiency | ≥87% (Energy star level "VI") |
Infrared radiation (light) | 850 nm, < 2mw/cm2 |
Operating Temperatures | 50°F to 104°F (+10°C to +40°C) |
Operating Humidity | 30% to 95% relative humidity (non-condensing) |
Operating Atmospheric Pressure | 800 hPA to 1060 hPA |
Storage/Shipping Temperatures | 32°F to 122°F (0°C to +50°C) |
Storage/Shipping Humidity | 10% to 95% relative humidity (non-condensing) |
Storage/Shipping Atmospheric Pressure | 700 hPA to 1060 hPA |
Spherical Equivalent | Range: -7.50D to +7.50D in 0.25D increments Accuracy: -3.50D to 3.50D, ± 0.50D; -7.50D to < -3.50D, ± 1.00D; > 3.50D to 7.50D, ± 1.00D |
Cylindrical | Range: -3.00D to +3.00D in 0.25D increments Accuracy: -1.50D to 1.50D, ± 0.50D; -3.00D to < -1.50D, ± 1.00D; > 1.50D to 3.00D, ± 1.00D |
Cylindrical Axis | Range: 1 to 180 degrees in 1 degree increments Accuracy: ± 10 degrees (for cylinder values > 0.5D) |
Pupil Size | 3.0 mm to 9.0 mm in 0.1 mm increments in ± cylinder* Accuracy: ± 0.4 mm |
Pupil Distance | Range: 35 mm to 80 mm in 1 mm increments Accuracy: ± 1.5 mm |
Gaze Measurement | 0 to 20 degrees in increments of 1 degree nasal, temporal, inferior and superior Accuracy: ± 1.5 degrees |
Time Per Measurement | Second(s) |
Measuring Distance | 3.3 ft ± 2 in (1 m ± 5 cm) |
Fixation Target | Random visual pattern and audible sound |
Measurement Principal | Optical reflex |