Welch Allyn TM286 AutoTymp Pure Tone and Tympanometry Screener with Tympanometry Mode, Reflex Mode, and Audiometry Mode; Audiometry Headset (External), Threshold Audiometry Card, Thermal Printer Paper (3 rolls), Ear-Tip Set (#26100); 100-240 V, 50-60 Hz AC, IEC Plug Type-B.
Features
Fast Detection Of Middle Ear Pathologies
226 Hz Tympanograms Offer Summary Information
Manual And Automated Audiometry
Lightweight And Portable
Internal Printer
Acoustic Reflex Measurements Available In 500, 1000, 2000, 4000 Hz Frequencies
The Welch Allyn TM 286 AutoTymp is a portable, lightweight instrument used for screening tympanometry, audiometry, and acoustic reflex measurements. It is designed for fast and accurate detection of middle ear pathologies, with options for probe tones and optional accessories.
Note:
The instrument is to be used by trained personnel only.
The instrument is Rx Only.
For the State of California:
Warning: This product can expose you to chemicals known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.
×
Rally #
45-1-2860
Manufacturer #
28600
GTIN
00732094082036
Brand
TM 286 Autotymp
Manufacturer
Baxter / Hillrom / Welch Allyn
Country of Origin
Unknown
Application
Tympanometer
UNSPSC Code
42182416
Dimensions
12.5" W x 14.5" D x 4.7" H (31.8 cm W x 36.8 cm D x 11.9 cm H)
Weight
5 lbs (2.3 kg) - unit and probe
Operating Temperature
59° F to 104° F (15° C to 40° C)
Warm-up Time
10 minutes for instruments stored at room temperature
Storage/Shipping
-30° F to 149° F (-34° C to 65° C)
Ambient Pressure
98 kPa to 104 kPa
Humidity
15% to 95%
Warranty
One Year Warranty
Probe Tone
226 Hz, ±2%
Sound Pressure Level
226 Hz: 85.5 dB SPL, ±2.0 dB, measured in a 2.0 cm3 coupler
Harmonic Distortion
< 3%
Admittance (Compliance)
Range: 226 Hz: 0.0 to 1.5 cm3 or 0.0 to 3.0 cm3
Note:
The range is automatically selected based upon the amplitude of the compensated tympanogram.
The maximum uncompensated (ECV + tympanogram peak) admittance (compliance) range is 0 to 5.0 cm3.
ECV/cavity limits for initiating pressurization is 0.2 to 5.0 cm3. Compliance Accuracy: ±0.1 cm 3 or ±5%, whichever is greater.
Pressure Range
+200 to -400 daPa
Pressure Accuracy
±10 daPa or ±15%, whichever is greater
Rate of Sweep
600 daPa/sec except near tympanogram peak where sweep rate slows to 200 daPa/sec to provide better definition of peak compliance.
Direction of Sweep
Positive to negative
Tympanogram Test Time
Approximately 1 second
Gradient
Measurement of the tympanogram width taken at 50% of peak compliance.
Note:
1 daPa = 1.02 mm H20
The pressure sweeps to at least -100 daPa. To save test time, pressure sweep stops when tympanogram returns to baseline after -100 daPa.
Full pressure sweep for 5 cm3 from sea level to 7000 ft. altitude with no leak.
High compliance tympanograms (test time) will take somewhat longer.
Frequencies
500, 1000, 2000, and 4000 Hz for ipsilateral stimulation
Accuracy
±3%
Total Harmonic Distortion
< 5% for outputs less than 110 dB HL and <10% at 110 dB HL
Rise/Fall Time
5 to 10 msec
Transducers
Ipsilateral
Probe Assembly
Output Levels
Ipsilateral
500 and 4000 Hz: 80, 90,100 dB HL 1000 and 2000 Hz: 85, 95, 105 dB HL
Pressure
Reflex measures are set automatically to pressure at peak compliance with an offset of -20 daPa if peak pressure is negative and +20 daPa if peak pressure is positive.
Reflex Determination
Compliance change of 0.05 cm3 or greater
Reflex Test Time
1 to 12 seconds depending upon the number of ipsilateral test frequencies selected (4 maximum) and intensity required.
Note:
Ipsilateral stimuli are time multiplexed with probe tone (93 ms ON, 66 ms OFF).
Stimuli are presented at lowest level first. If there is no response, the intensity is increased by 10 dB until a response is detected or the maximum dB HL is reached.
Gloves Exam ProAdvantage® Medium NonSterile Latex Standard Cuff Length Textured Fingertips Ivory Not Chemo Approved (100/BX 10BXS/CS) by Pro Advantage ID P359103